◆ Software as a Medical Device

IEC 62304 Consulting Support

IEC 62304 lifecycle implementation, software documentation and verification & validation for Software as a Medical Device across FDA, EU MDR, CDSCO and 50+ markets.

IEC 62304 Software Lifecycle

IEC 62304 Software Lifecycle Implementation for SaMD

IEC 62304 is the international standard that defines the software lifecycle processes for medical device software, including Software as a Medical Device (SaMD). It requires a documented development plan, requirements, architecture, verification, risk management, configuration management, problem resolution and maintenance, scaled to the software’s safety class (A, B or C). The FDA recognises IEC 62304 as a consensus standard, EU MDR notified bodies treat it as the state of the art, and CDSCO expects it for software devices in India.

Most SaMD submissions that get deficiency letters are not rejected for the algorithm. They are rejected because the software lifecycle evidence is missing, inconsistent or reconstructed after the fact. NexorTest builds IEC 62304 into your engineering workflow from sprint one, so the Design History File and technical documentation come out of the work your team already does.

IEC 62304 clauseProcessWhat we implement and document
Clause 4General requirementsQMS link to ISO 13485, ISO 14971 risk process, software safety classification rationale, legacy software assessment (4.4)
5.1Software development planningSoftware Development Plan (SDP) covering lifecycle model, deliverables, traceability, tools, standards, and SOUP management
5.2Software requirements analysisSoftware Requirements Specification (SRS) with functional, performance, security, usability and risk-control requirements
5.3 – 5.4Architectural & detailed designSoftware architecture, segregation for risk control, SOUP identification, detailed unit design for Class C
5.5 – 5.7Unit, integration & system testingUnit verification criteria, integration test plan, system test protocols and reports, regression strategy
5.8Software releaseRelease checklist, known residual anomalies evaluation, version archive, reproducible build records
Clause 6Software maintenanceMaintenance plan, change impact analysis, post-market feedback loop, PCCP alignment for AI/ML updates
Clause 7Software risk managementSoftware hazard analysis, contributing causes, risk controls traced to requirements and tests (ISO 14971, IEC/TR 80002-1)
Clause 8Configuration managementConfiguration items, branching and baseline strategy, SOUP and dependency version control, SBOM
Clause 9Problem resolutionProblem report workflow, trend analysis, link to software CAPA and vigilance

Compliance basis: IEC 62304:2006 + Amendment 1:2015 (EN 62304:2006/A1:2015), the edition currently recognised by FDA and assessed by EU notified bodies.

Software Safety Classes

IEC 62304 Software Safety Classification: Class A, B and C

IEC 62304 assigns each software system a safety class based on the worst harm a software failure could contribute to: Class A means no injury is possible, Class B means non-serious injury is possible, and Class C means death or serious injury is possible. Software failure is assumed to happen (100% probability), so only risk controls outside the software can lower the class. The higher the class, the more documentation and testing the standard requires.

Deliverable / activityClass AClass BClass C
Software development plan✓✓✓
Software requirements specification✓✓✓
Software architectureNot required✓✓
Detailed design of each software unitNot requiredNot required✓
Unit verification (with acceptance criteria)Not required✓✓ + additional criteria
Integration testingNot required✓✓
Software system testing✓✓✓
Software risk management (Clause 7)Partial✓✓
SOUP requirements & known anomaly evaluationNot required✓✓
Release, maintenance, configuration & problem resolution✓✓✓

Simplified summary. Exact sub-clause applicability is defined in IEC 62304:2006+A1:2015. Most diagnostic, triage and AI/ML SaMD lands in Class B or C.

Verification & Validation

SaMD Verification and Validation (V&V) Studies

Verification proves the software was built right: each requirement is implemented and tested at unit, integration and system level. Validation proves the right software was built: the finished SaMD meets user needs and its intended use in the real clinical environment, through usability, analytical and clinical performance studies. Regulators expect both, linked by a traceability matrix from user needs to requirements, risks and test results.

🧪
Software verification testing
🩺
Design and clinical validation
🔒
Cybersecurity and non-functional testing
Traceability that survives an audit

Every requirement, risk control and test is linked in one traceability matrix: user need → system requirement → software requirement → architecture item → risk control → verification test → validation result. Reviewers can follow any line in both directions, and every change triggers an impact analysis rather than a re-write of the file.

AI/ML model validation

For machine learning SaMD we document training, tuning and test dataset separation, reference standard definition, subgroup and bias analysis, multi-site performance, and the verification protocol your PCCP will use for future model updates.

Submission ready documentation

IEC 62304 Documentation for FDA, EU MDR and CDSCO Submissions

One IEC 62304 documentation set can support FDA, EU MDR and CDSCO submissions. The FDA’s Content of Premarket Submissions for Device Software Functions guidance (2023) asks for Basic or Enhanced documentation depending on risk. EU MDR Annex I §17.2 requires software to follow a state-of-the-art lifecycle with risk management, information security, verification and validation. CDSCO requires the same evidence in the device master file.

DocumentFDA BasicFDA EnhancedEU MDR technical documentationCDSCO (India)
Documentation level evaluation / software description✓✓✓✓
Risk management file (ISO 14971)✓✓✓✓
Software requirements specification✓✓✓✓
System and software architecture diagram✓✓✓✓
Software design specificationNot required✓✓✓
Lifecycle evidence (IEC 62304 declaration or plans)Declaration of conformity or summaryPlus configuration & maintenance plans✓✓
Software testing (V&V)System test summary + traceabilityPlus unit & integration test evidence✓✓
Version history and unresolved anomalies✓✓✓✓
Cybersecurity documentation and SBOM✓✓✓✓

Also applies to UKCA (MHRA), TGA, Health Canada, SFDA and other IMDRF-aligned regulators. Cybersecurity content for FDA follows section 524B of the FD&C Act for cyber devices.

How we work

Our IEC 62304 Implementation Process

Whether you are a start-up writing your first line of code or a scale-up with a product already in the market, we take you from where you are to a submission-ready software file.

1
IEC 62304 gap assessment

We review your code repository, tools, tickets and existing documents against every applicable IEC 62304 clause and the FDA software guidance, and give you a prioritised remediation plan.

2
Safety classification and development plan

We determine the software safety class with a documented rationale, define software items, and write the Software Development Plan, SOUP list and configuration management plan.

3
Agile-compatible lifecycle set-up

We map IEC 62304 activities onto your sprints in Jira, Azure DevOps, GitHub or GitLab (following AAMI TIR45), so requirements, risks and tests are traced automatically instead of in spreadsheets.

4
Design documentation and software risk management

We produce the SRS, architecture, detailed design (Class C) and software hazard analysis, with every risk control linked to a requirement and a test.

5
Verification and validation execution

We write the V&V protocols, execute or witness testing, run usability and cybersecurity testing through our labs, and compile test reports and the traceability matrix.

6
Submission, release and maintenance

We compile the FDA eSTAR software section, EU MDR technical documentation and CDSCO dossier, then keep the file current through releases, patches and AI/ML model updates.

Regulatory update

IEC 62304 Edition 2: What Is Changing for SaMD?

IEC 62304 Edition 2 is being developed to replace the 2006 standard and its 2015 amendment. Drafts replace the three safety classes (A, B, C) with two process rigor levels, extend the scope to all health software, and add explicit expectations for AI/ML and cybersecurity. Publication has been delayed several times. Until it is published and recognised by FDA and the EU, IEC 62304:2006+A1:2015 remains the version your submissions are assessed against.

How we prepare you

We build your lifecycle to the current edition, with the extra rigor that Edition 2 drafts point towards, so moving to the new edition does not mean rewriting your Design History File. Regulators are expected to allow a transition period after publication.

IEC 62304 FAQ

IEC 62304 FAQ

IEC 62304 is not written into law, but in practice it is required. The FDA recognises it as a consensus standard and expects lifecycle evidence in every software submission. EU MDR Annex I §17.2 requires state-of-the-art software lifecycle processes, which notified bodies assess against EN 62304. CDSCO, MHRA, TGA and Health Canada expect the same. Not following IEC 62304 means proving an equivalent process, which is harder and slower.

Depending on the safety class: a software development plan, software requirements specification, software architecture, detailed design (Class C), unit, integration and system test records, software risk management file, SOUP list, configuration management records, problem reports, release records and a maintenance plan. All of them must be traceable to each other.

Assume the software fails, then identify the worst harm that failure could contribute to after risk controls external to the software. No injury possible is Class A, non-serious injury is Class B, and serious injury or death is Class C. The rationale must be documented in the risk management file.

Verification confirms the software meets its specified requirements (unit, integration and system testing). Validation confirms the finished SaMD meets user needs and its intended use, through usability testing, analytical validation and clinical performance evaluation. Both are required for FDA clearance and CE marking.

SOUP (Software of Unknown Provenance) is third-party or open-source software not developed under IEC 62304, such as operating systems, libraries and frameworks. You must list each SOUP item with its version, define its functional and performance requirements, specify the hardware and software it needs, and evaluate its published anomalies for risk.

IEC 62304 covers the software development lifecycle. IEC 82304-1 covers the whole standalone health software product: product requirements, product validation, accompanying documents and post-market activities. Standalone SaMD usually needs both, with IEC 82304-1 referencing IEC 62304 for the development process.

IEC 62304 is not a management-system standard like ISO 13485, so there is no mandatory IEC 62304 certificate. Conformity is shown in your FDA submission or EU technical documentation and checked during audits. Some test bodies offer voluntary IEC 62304 assessments, which NexorTest can prepare you for.

Yes. AI/ML SaMD follows IEC 62304 for the software around the model, while the model itself needs additional evidence: dataset management, training and test separation, bias analysis and performance validation. For FDA, algorithm changes after clearance are managed through a Predetermined Change Control Plan (PCCP).

Ready to Get a Free IEC 62304 Gap Assessment?

Our specialists review your code repository, tools, tickets and existing documents against every applicable IEC 62304 clause and the FDA software guidance, then give you a prioritised remediation plan for FDA, EU MDR and CDSCO submissions.