IEC 62304 Consulting Support
IEC 62304 lifecycle implementation, software documentation and verification & validation for Software as a Medical Device across FDA, EU MDR, CDSCO and 50+ markets.
IEC 62304 Software Lifecycle Implementation for SaMD
IEC 62304 is the international standard that defines the software lifecycle processes for medical device software, including Software as a Medical Device (SaMD). It requires a documented development plan, requirements, architecture, verification, risk management, configuration management, problem resolution and maintenance, scaled to the software’s safety class (A, B or C). The FDA recognises IEC 62304 as a consensus standard, EU MDR notified bodies treat it as the state of the art, and CDSCO expects it for software devices in India.
In our experience, most FDA requests for additional information on SaMD concern documentation (traceability, anomalies, SOUP, cybersecurity) rather than algorithm performance. They are rejected because the software lifecycle evidence is missing, inconsistent or reconstructed after the fact. NexorTest builds IEC 62304 into your engineering workflow from sprint one, so the Design History File and technical documentation come out of the work your team already does.
| IEC 62304 clause | Process | What we implement and document |
|---|---|---|
| Clause 4 | General requirements | QMS link to ISO 13485 and the US QMSR (ISO 13485 incorporated into FDA rules from Feb 2026), ISO 14971 risk process, software safety classification rationale, legacy software assessment (4.4) |
| 5.1 | Software development planning | Software Development Plan (SDP) covering lifecycle model, deliverables, traceability, tools, standards, and SOUP management; control of supporting tools (5.1.10), configuration items placed under control before verification (5.1.11), and identification of common software defect categories (5.1.12, e.g. OWASP Top 10 and CWE Top 25 for web and cloud SaMD) |
| 5.2 | Software requirements analysis | Software Requirements Specification (SRS) with functional, performance, security, usability and risk-control requirements |
| 5.3 – 5.4 | Architectural & detailed design | Software architecture, segregation for risk control (a software item can carry a lower class than the system only when its segregation is documented with a rationale, 5.3.5), SOUP identification, detailed unit design for Class C |
| 5.5 – 5.7 | Unit, integration & system testing | Unit verification criteria, integration test plan, system test protocols and reports, regression strategy |
| 5.8 | Software release | Release checklist, known residual anomalies evaluation, version archive, reproducible build records |
| Clause 6 | Software maintenance | Maintenance plan, change impact analysis, post-market feedback loop, PCCP alignment for AI/ML updates |
| Clause 7 | Software risk management | Software hazard analysis, contributing causes, risk controls traced to requirements and tests (ISO 14971, IEC/TR 80002-1); risk management of software changes, including SOUP changes (7.4) |
| Clause 8 | Configuration management | Configuration items, branching and baseline strategy, SOUP and dependency version control. The SBOM is maintained alongside, as an FDA cybersecurity requirement for cyber devices (FD&C Act 524B) rather than an IEC 62304 requirement |
| Clause 9 | Problem resolution | Problem report workflow, trend analysis, link to software CAPA and vigilance |
Compliance basis: IEC 62304:2006 + Amendment 1:2015 (EN 62304:2006/A1:2015), the edition currently recognised by FDA and assessed by EU notified bodies.
IEC 62304 Software Safety Classification: Class A, B and C
IEC 62304 assigns each software system a safety class based on the worst harm a software failure could contribute to: Class A – no injury or damage to health is possible; B – non-serious injury possible; C – death or serious injury possible. Software failure is assumed to happen (100% probability), so only risk controls outside the software can lower the class. The higher the class, the more documentation and testing the standard requires.
Classifying parts of the software. Each software item inherits the safety class of the software system it belongs to. A software item can be given a lower class only when it is segregated from higher-class items and the segregation, with its rationale, is documented in the software architecture (IEC 62304 cl. 4.3 d-f and 5.3.5). Without that documented segregation, the whole system is treated at its highest class.
| Deliverable / activity | Class A | Class B | Class C |
|---|---|---|---|
| Software development plan | ✓ | ✓ | ✓ |
| Software requirements specification | ✓ | ✓ | ✓ |
| Software architecture | Not required | ✓ | ✓ |
| Software units identified (5.4.1) | Not required | ✓ | ✓ |
| Detailed design of each software unit | Not required | Not required | ✓ |
| Unit verification (with acceptance criteria) | Not required | ✓ | ✓ + additional criteria |
| Integration testing | Not required | ✓ | ✓ |
| Software system testing | ✓ | ✓ | ✓ |
| Software risk management (Clause 7) | Partial | ✓ | ✓ |
| SOUP requirements & known anomaly evaluation | ✓ | ✓ | ✓ |
| Release, maintenance, configuration & problem resolution | Not required | ✓ | ✓ |
Simplified summary. Exact sub-clause applicability is defined in IEC 62304:2006+A1:2015. Most diagnostic, triage and AI/ML SaMD lands in Class B or C.
Worked example: a cloud tele-ophthalmology / image-management SaMD was reclassified from Class A to Class B. “A physician always reviews” is not an external risk control when the physician cannot detect the failure, for example a wrong-patient image.
SaMD Verification and Validation (V&V) Studies
Verification proves the software was built right: each requirement is implemented and tested at unit, integration and system level. Validation proves the right software was built: the finished SaMD meets user needs and its intended use in the real clinical environment, through usability, analytical and clinical performance studies. Regulators expect both, linked by a traceability matrix from user needs to requirements, risks and test results.
Software verification testing
- Software V&V plan and test strategy
- Unit test criteria and code review records
- Integration and system test protocols and reports
- SOUP and third-party library verification
- Automated regression suites for CI/CD pipelines
- Unresolved anomaly list with risk justification
Design and clinical validation
- Usability engineering and summative validation (IEC 62366-1)
- Health software product validation (IEC 82304-1)
- Analytical validation: sensitivity, specificity, AUC, reproducibility
- Clinical performance and clinical evaluation support
- Standalone and reader studies for AI/ML imaging
Cybersecurity and non-functional testing
- Penetration testing and vulnerability scanning (IEC 81001-5-1)
- SBOM and SOUP vulnerability monitoring
- Threat model
- Security architecture views
- Post-release vulnerability management plan
- Post-remediation retest evidence
- Performance, load and interoperability testing
- Platform and OS compatibility for mobile and cloud SaMD
- Test evidence formatted for FDA and notified body review
Traceability that survives an audit
Every requirement, risk control and test is linked in one traceability matrix: user need → system requirement → software requirement → architecture item → risk control → verification test → validation result. Reviewers can follow any line in both directions, and every change triggers an impact analysis rather than a re-write of the file.
AI/ML model validation
For machine learning SaMD we document training, tuning and test dataset separation, reference standard definition, subgroup and bias analysis, multi-site performance, and the verification protocol your PCCP will use for future model updates.
IEC 62304 Documentation for FDA, EU MDR and CDSCO Submissions
One IEC 62304 documentation set can support FDA, EU MDR and CDSCO submissions. The FDA’s Content of Premarket Submissions for Device Software Functions guidance (2023) asks for Basic or Enhanced documentation depending on risk. EU MDR Annex I §17.2 requires software to follow a state-of-the-art lifecycle with risk management, information security, verification and validation. CDSCO requires the same evidence in the device master file.
Still using Level of Concern? The 2023 guidance replaced the Minor, Moderate and Major Level of Concern from the FDA’s 2005 software guidance with two documentation levels. Broadly, software that was Major Level of Concern now needs Enhanced documentation, and Minor or Moderate usually maps to Basic. The level must be justified again against the 2023 criteria rather than carried over from old SOPs.
| Document | FDA Basic | FDA Enhanced | EU MDR technical documentation | CDSCO (India) |
|---|---|---|---|---|
| Documentation level evaluation / software description | ✓ | ✓ | ✓ | ✓ |
| Risk management file (ISO 14971) | ✓ | ✓ | ✓ | ✓ |
| Software requirements specification | ✓ | ✓ | ✓ | ✓ |
| System and software architecture diagram | ✓ | ✓ | ✓ | ✓ |
| Software design specification | Not required | ✓ | ✓ | ✓ |
| Lifecycle evidence (IEC 62304 declaration or plans) | Declaration of conformity or summary | Plus configuration & maintenance plans | ✓ | ✓ |
| Software testing (V&V) | System-level test protocol and report (expected and actual results, pass/fail) + traceability | Plus unit & integration test evidence | ✓ | ✓ |
| Version history and unresolved anomalies | ✓ | ✓ | ✓ | ✓ |
| Cybersecurity documentation and SBOM | For cyber devices / where applicable | ✓ | ✓ | ✓ |
Also applies to UKCA (MHRA), TGA, Health Canada, SFDA and other IMDRF aligned regulators. Cybersecurity content for FDA follows section 524B of the FD&C Act for cyber devices.
Our IEC 62304 Implementation Process
Whether you are a start-up writing your first line of code or a scale-up with a product already in the market (including legacy software that needs remediation), we take you from where you are to a submission-ready software file.
IEC 62304 gap assessment
We review your code repository, tools, tickets and existing documents against every applicable IEC 62304 clause and the FDA software guidance, and give you a prioritised remediation plan. For software already on the market, our legacy and remediation track covers the IEC 62304 4.4 gap analysis, retrospective documentation, reclassification after reviewer pushback, and reconciling inconsistent records (SBOM against the SOUP list, IFU statements against verified features).
Safety classification and development plan
We determine the software safety class with a documented rationale, define software items, and write the Software Development Plan, SOUP list and configuration management plan.
Agile-compatible lifecycle set-up
We map IEC 62304 activities onto your sprints in Jira, Azure DevOps, GitHub or GitLab (following AAMI TIR45), so requirements, risks and tests are traced automatically instead of in spreadsheets. We also validate these tools (code repository, CI/CD, test tools, static analysis and dependency scanners).
Design documentation and software risk management
We produce the SRS, architecture, detailed design (Class C) and software hazard analysis, with every risk control linked to a requirement and a test. We add design reviews with independent reviewer, software design history file, and the US QMSR (ISO 13485 incorporated into FDA rules from Feb 2026).
Verification and validation execution
We write the V&V protocols, execute or witness testing, run usability and cybersecurity testing through our labs, and compile test reports and the traceability matrix.
Submission, release and maintenance
We compile the FDA eSTAR software section, EU MDR technical documentation and CDSCO dossier, then keep the file current through releases, patches and AI/ML model updates. FDA guidance on when a software change needs a new 510(k), and letter-to-file practice.
IEC 62304 Edition 2: What Is Changing for SaMD?
As of October 2026, IEC 62304 Edition 2 is being developed to replace the 2006 standard and its 2015 amendment. Drafts replace the three safety classes (A, B, C) with two process rigor levels, extend the scope to all health software, and add explicit expectations for AI/ML and cybersecurity. Publication has been delayed several times. Until it is published and recognised by FDA and the EU, IEC 62304:2006+A1:2015 remains the version your submissions are assessed against.
- Process rigor levels: Classes A/B/C replaced by process rigor levels; today's Class B software is expected to need the higher rigor level
- Broader scope: Scope broadened from medical device software to health software
- Security and AI/ML: Stronger links to cybersecurity (IEC 81001-5-1) and AI/ML development practices
- IEC project page: IEC 62304 Ed. 2, Health software Software life cycle processes
How we prepare you
We build your lifecycle to the current edition, with the extra rigor that Edition 2 drafts point towards, so moving to the new edition does not mean rewriting your Design History File. Regulators are expected to allow a transition period after publication.
Frequently Asked Questions on IEC 62304
Can the IEC 62304 safety class be lowered?
Yes, but only through a risk control outside the software system (for example a hardware interlock or an independent check) or by segregating a software item and documenting the rationale. Arguments that the software rarely fails do not count, because IEC 62304 assumes the software will fail. If a reviewer rejects the rationale, expect to reclassify and add the missing Class B or C deliverables.
Can a clinician's review count as a risk control?
Only if the clinician can actually detect the failure in time to prevent harm. A clinician can catch an implausible result, but not an image filed under the wrong patient, a silently dropped finding or a corrupted value that still looks normal. If the failure cannot be detected, the review is not an effective external risk control and the higher class stays.
Does IEC 62304 cover cloud infrastructure?
Partly. IEC 62304 covers the software you develop and the SOUP it relies on, including cloud services and platform components treated as SOUP. The hosting infrastructure itself is not an IEC 62304 deliverable, but it must be addressed through risk management, cybersecurity (IEC 81001-5-1 and the FDA premarket cybersecurity guidance) and supplier controls under ISO 13485.
Is an SBOM part of IEC 62304?
No. IEC 62304 requires a SOUP list (title, manufacturer and version for each item) and an evaluation of SOUP anomalies, but not an SBOM. The FDA requires an SBOM for cyber devices under section 524B of the FD&C Act. In practice the SBOM and the SOUP register should come from the same source and stay consistent.
Do software development tools need validating?
Yes. ISO 13485 (clauses 4.1.6 and 7.5.6) requires validation of software used in the quality system and in production, proportionate to risk, and IEC 62304 (5.1.10) requires planning how supporting tools are controlled. This typically covers the issue tracker, code repository, CI/CD pipeline, test automation, static analysis and dependency scanners. Validation focuses on how you use each tool, not on retesting the vendor's product.
Ready to Get a Free IEC 62304 Gap Assessment?
Our specialists review your code repository, tools, tickets and existing documents against every applicable IEC 62304 clause and the FDA software guidance, then give you a prioritised remediation plan for FDA, EU MDR and CDSCO submissions.